The unique in-line pressure-sensing system provides instantaneous adjustments to flow levels based on intraarticular pressure changes,
providing ideal joint distention throughout the procedure. Pressure and flow levels can be adjusted independently on the pump or from
the sterile field with the autoclavable remote for optimum control of fluid dynamics. Clear joint distention is maintained even with
aggressive shaving, aspiration and multiple arthroscopy portals.
Specifications
Width 14.5 inches (36.5 cm) Height 5.0 inches (12.5 cm) Depth 12 inches (30 cm)
Weight 18 pounds (8.2 kg)
Maximum Flow rate 1500 ml/minute
Pressure 10 – 120 mmHg increments of 5 mmHg.
Overpressure control 300 mmHg
Pressure control Continuous pressure checking
Protection IPX1
Main cable 10 A/250 V
Connector CEE 7/7
Jack IEC 320/C13
Power supply 100-240 V, 50/60 Hz, 6.5A
Mains Fuse T2.0A 250V
Cleaning Surface cleaning with mild detergent sterilization Surface disinfection with mild disinfectant