Zimmer ATS 2000

Is a dual pump system that allows for Bier Block procedures. Feel free to inquire about Zimmer, Scanmed, and Expressaire tourniquet cuffs and hoses. Single & Dual Port Options: All cuff styles are available in both single and dual port configurations, allowing for use with any A.T.S. tourniquet regulator. Single port cuffs for use with the A.T.S. 750 and A.T.S. 1200 Tourniquet Systems. Dual port and Bier Block cuffs for use with A.T.S. 1500 and 2000 Tourniquet System.

    • Automatic, microprocessor-based pneumatic tourniquet system.
    • Able to independently control and monitor two cuffs in a Bier Block or bilateral procedure.
    • Portable and designed for tabletop or tourniquet stand monitoring.

Specifications

Line Voltage Range: 90–240 V~, 50/60 Hz, auto switching

Line Current: 670 mA RMS @ 120 V~

Input Power: 53 Watts Typical

AC Indicator Light: Green LED

Battery Type: 2 Rechargeable, 12 V sealed lead acid, 2.3 Amp hours

Battery Discharge Time: Unit will operate on battery power for 60 minutes minimum with fully charged batteries.

Battery Recharge Time: 24 hours (Maximum) Unit should be plugged in 24 hours before initial use.

Power Cord: Type SJT, AWG 16, 14 ft. (4.27 m)

Power Plug: Hospital grade, 3 prong straight blade, 15 Amp

Line Protection: 2 time delayed 1.0 Amp 250 volt fuses

Cuff Pressure Range:  50–475 mm Hg

Pressure Accuracy: ±3 mm Hg (50–475 mm Hg)

Pressure Regulation: ±4 mm Hg of set point (10 second average under non-transient conditions without external leaks)

Maximum Pressures: 475 mm Hg cuffs700 mm Hg reservoir

Time Alarm Set Ranges: 0–240 minutes; 1 minute increments

Timer Accuracy: 0.25% ±1 second

Internal Diagnostics: Program, memory, watcHDog timer, transducer calibration, improper valve actuation.

Height: 12.75 in. (32.4 cm)

Width: 10 in. (25.4 cm)

Depth: 10.5 in. (26.7 cm) (including clamp)

Weight: 19.8 lbs (9 kg)